Technical Parameters
- Display Format : LED. SP02:35-99% in two-digit 7-segment
digital display. Pulse Ratio: 30-240 BPM in 3digit 7 segment digital
display Pulse Wave: 6-segment bar graph display. Battery Voltage Low
Indication: LED Signals.
- Power Requirement : Two AAA Size l.SV Alkali Battery.
- 3. Power Consumption : Less than 40mA
- Resolution's% for SPO2 and lBPM for Pulse Ratio.
- Measurements Accuracy : ±2% in stage of 70% - 99%. ±2
bpm for Pulse Ratio.
- Measurement Performance in weak filling condition. Oxyhemoglobin and
pulse value can be shown correctly when pulse -iilling ratio is 0.5%.
- Interference Resistance Capacity against Ambient Light :
Deviation is smaller than±1% between value of Oxyhemoglobin
measured in natural lighting indoor condition and present lighting
sources and that measured in dark room.
- It is equipped with a function switch, through which the Oximeter
can be automatically off when no finger is in the Oximeter more then 8
Seconds.
.
Product Properties
- Operation of the product is simple and convenient
- The product is small in volume, light in weight (total weight is
about SOg including batteries) and convenient in carrying.
- Power Consumption of the product is low and the two
originally-equipped two AAA batteries can be operated continuously for
30 hours.
- Low voltage warning will be indicated in visual window when battery
voltage is too low then normal operation of the Oximeter might be
influenced.
- The product will automatically be powered off when no signal is in
the product for more then 8 seconds.
.
Product Operation Scope
The fingertip Oximeter can be used to measure human
Hemoglobin Saturation and heart beat through finger. The product is suitable
for use in family, hospital (including clinical use in
internist/surgery,Anesthesia,padiatrics, intensive care and etc) Oxygen Bar,
Social medical organizations ,physical care in sports (it can be used before
or after sports). Continually monitoring is not recommended.
Specification
Integrated with SPO2 probe and processing display
module Can be used in: hospital, home, community medical treatment, sports
healthcare, etc. Can measure SPO2 and PR accurately SPO2 and PR display, PR
waveform and bargraph display Battery voltage indicator Friendly Operation
Menu for The Function Setting 4 Direction Dispaly Four levels adjustable
brightness Pulse sound function Alarm function; can set upper and lower
limits. Can store data USB interface to Computer, software can analyse,
store and print the data. Rechargeablr Li battery Low power consumption;
shut off automatically when no signal Small volume(57(L) ¡Á
32(W) ¡Á30(H) mm)¡¢light£convenient to carry .
Technical Parameters
1.3' 65K Color OLED display, resolution 160*128
SPO2 : 35¡"99£¥ Pulse Ratio£º30¡"250BPM
Resolution : 1% for SPO2, 1BPM for pulse ratio
Accuracy : ¡À2% (70£¥¡"99£¥),
unspecified (<70%) for SPO2, À2BPM or¡À 2% (select
larger) for pulse ratio
Power : 1 rechargeable Li battery
Charge manner : AC transformer or PC USB port Can work more than 24
hours continuously Interference resistance capacity against ambient light
and measurement performance at low perfusion .
Specification
Enertech's E-300 generators gives consistent and
repeatable output as required by the surgeon in different types of
surgeries.
E-300 electro surgery generators gives best price performance ratio in
similar class of generators offered by other manufacturers and can be widely
used in procedures including Urology, GI, Gynaecologic, Orthopaedic,
Thoracic and Plastic Surgery.
Intensive Care Warmer
- With baby care trolley
- with utility tray top.
- provision for one oxygen cylinder
- Provided with I.V. rod.
- Special spot light.
- with large baby basinet.
- with electronic temperature indicator.
Easy To Use
As technology advanced, we have kept the Warmer
simple & easy to use with one touch controls.
Focused Heating
The heater with reflector makes it easier to work
with the baby by efficiently providing even heat distribution across the
cradle and not into the surrounding environment.
Servo Control Mode
Servo Control is an electronic feedback system which
functions as a thermostat to maintain a constant temperature at the site of
Skin Probe by regulation the heat output of Radiant Warmer Maintaining a
constant abdominal Skin temperature between 36.5 & 37.0 C is the
simplest way to provide a "thermoneutral" environment, minimizing
the no. of calories needed to maintain normal body temperature &
reducing the risks of cold Stress or overheating. Servo control is the only
acceptable method of heat regulation for the Infant cared for under a
Radiant Warmer.
Skin Control Mode
Skin Control is an electronic feedback system which
functions as a thermostat to maintain a constant temperature of baby cradle
by regulating the heat output of Radiant Warmer. It also acts as a WATCH DOG
as an Air over temperature cuts off heater even if the Warmer is operating
in Servo mode.
Manual Mode
The manual mode allows the Warmer to be kept 'baby
ready' without nuisance alarms. When needed the Warmer provides safe,
powerful & controlled warming to restore the temperature drop
experienced by the baby immediately after delivery.
Mechanical Specification
- Easy access of Infant from all sides with thick Acrylic collapsible
side panels
- Acrylic baby tray with Infant's head up / down facility &foam
mattress.
- Provision for positioning X-ray cassette
- S. S. tray with a lot to use under surface Phototherapy
- Swiveling Warming source for taking X-ray
- Consist of CERAMIC INFRAN RED / SILICON QUARTZ GLASS HEATER
- S. S. Reflector with safety grill designed such to achieve uniform
temperature & easy cleaning without any electrical shock hazard
- Instrument tray, pole for Infusion pump & I. V. Stand
- Examination lamp with sufficient illumination to accurate infant
color during procedures
- M. S. Fabrication with powder coated paint
- Mounted on castors for easy mobility
Technical Specifications
 |
| Power Source |
230V AC, +- 10%, 50 HZ |
| Light Source |
60 Watt Bulb to examine babies |
| Heater Capacity |
600 Watts |
| Power Consumption |
800 Watts at 100% Heater power |
|
 |
Neocare Warmer (supplied with skin probe)
- Skin Temperature displayed individually on 1" bright Red LED
display
- Set Temperature displayed on separate bright Green LED display
- Two modes of Warming i. e. Skin, & Manual
- Heater output in 10 equal steps
- Programmable heater power & Time in manual mode
- Stain free read alarm messages
- Individual LED for selection of mode (Skin, Air & Manual)
Infant Warmer (Skin /Manual)
- Skin Temperature displayed on 1" bright Red LED display
- Set Temperature displayed separate bright Green LED display
- Two modes of Warming i. e. Skin & Manual
- Heater output in 10 equal steps
- Programmable Heater power & Time in Manual mode
- Stain free read alarm messages
- Individual LED for selection of mode (Skin & manual)