Specifications
- LCD Based Pulse Oxymeter.
- Table Top Model.
- Response Mode: Adult, Pediatric & Neonate.
- Continuous Plethysmograph Display & Bar Graph.
- Pulse Rate: 20 250 BPM.
- Alarm Setting for SPO2 & Pulse Rate High & Low.
- Displayed Trends 24 Hr - Spo2% Pulse Rate & Status.
- Real Time & Date Display.
- Alarm Visual & Audio.
- Screen Size 110 mm x 85 mm & More.
- Battery Backup 3 Hours.
- Set up Menu: Predefined Parameter Selection.
- A.C. Mains Variation : 100-260 V AC
- Power Consumption : <20 VA.
- Ratient Safety: Class 1 BF - ISO Standard.
Other Additional Specifications
Table Top Pulse Oximeter Product Description
Pulse Rate:
Protection:
- Up to 450 records storage
- Up to 12 hours of trend display
- Built in rechargeable Li battery
- Adult, Pediatric application
- Pleth waveform display
.
Audible and Visual alarm SpO2 :
 |
| Range |
0 to 100% |
| Resolution |
1% |
| Accuracy |
+/-2% (70 to 100%) |
| Average mode |
8 pulse beat |
|
 |
Pulse Rate
 |
| Range |
30 to 250bpm |
| Resolution |
1bpm |
| Accuracy |
+/-2bpm(30 to 250bpm) |
| Average mode |
1 pulse beat |
|
 |
Protection
Withstand 4500VAC/50Hz/60Hz Voltage in isolation
against Electrosurgical interference And defibrillation
Pulse Oximeter Vital Signs
Specification
Monitor
compact and portable, allowing for uninterrupted
monitoring with built-in battery high bright LEDS display of NIBP,SPO2 and
pulse rate high resolution color LCD for trend tabular and SPO2 waveforms
display manual/auto/continuous measurement of NIBP adjustable audible and
visual alarms Up to 600 groups NIBP data up to 10 hours suitable for adult,
pediatric and neonatal patient
Display
7 segment LEDS for systolic,diastolic,mean
pressures,SPO2 and pulse rate 2.4``color LCD for waveforms,bargraphic,trend
and system menu
NIBP
 |
| Method |
Oscillometry |
| Operation Modes |
Manual/Automatic/STAT |
| Auto Measure Time |
5~250 minutes adjustable, interval 5 minutes
|
| Measurement Range |
10~270mmHg |
| Overpressure Protection |
yes |
| Resolution |
1mmHg |
| Alarm |
systolic,Diastolic,Mean |
|
 |
SPO2
 |
| Measurement Range |
0~100% |
| Resolution |
1% |
| Accuracy |
70~100% ¡À2%,0~69% unspecified |
| Alarm Range |
0~100% |
|
 |
Pulse Rate
 |
| Range |
0~254bpm |
| Resolution |
1bpm |
| Accuracy |
¡À1bpm |
| Alarm Range |
0~100% |
|
 |
Option
 |
| ECG,Thermal printer |
| Dimension |
200*230*75mm |
| Weight |
about 3.5kg |
| Power Supply |
AC 100~250V 50/60Hz |
|
 |
Cardiac Monitor Multi-parameter Monitor
Features
- 8.4¡å color TFT display
- Suitable for adult, pediatric and neonatal patient
- Basic parameters (ECG, SpO2, NIBP) in a durable case for bedside
monitoring and transport
- Audible and visual alarms with adjustable alarm ranges
- Networkable with CONTEC central monitoring system
- Powerful data management and storage capacity
ECG
 |
| Lead type |
5-lead |
| Input |
RA; LA; RL; LL; V |
| Sweep speed |
12.5mm/s, 25mm/s, 50mm/s |
| Protection |
Withstand 4000VAC/50Hz voltage in isolation
against electrosurgical and defibrillation |
| S-T detection |
YES Arrhythmia analysis: YES |
| Alarm |
YES, audible and visual alarm, alarm events
recallable |
|
 |
NIBP
 |
| Method |
Oscillometry |
| Operation Modes |
Manual/Automatic/STAT |
| Measurement Unit |
mmhg/kpa selectable |
| Measurement Types |
Systolic, Diastolic, Mean |
| Over-Pressure Protection |
YES |
|
 |
SpO2
 |
| Measurement Range |
0-100% |
| Accuracy |
digit (70~100%); 0~69% unspecified |
|
 |
Standard Configuration
ECG, NIBP, SpO2
Optional
RESP,2-TEMP,2-IBP,ETCO2,Network,Thermal Recorder
Cardiac Monitor-Multi-parameter Monitor
Item specification
- Having facility to monitor ECG, SpO2, NIBP
- Integrated high resolution LCD screen.
- Size of display 12.1 inches.
- Waveform display moving and fixed form.
- Having arrhythmia detection with alarm facility.
- Digital alphanumeric key board to maintain patient name and ID.
- Direct function keys for frequently used monitor functions like -
print and BP
- Digital contract adjustment for clear viewing.
- Regular display of selected ECG leads and heart rate.
- Operate on mains & battery.
- Battery backup minimum 4-6 hours approx.
- Alarm limits with alarm levels and indicators.
- Can be upgradeable to - IBP, ET CO2 & internal thermal printer.
- 5 lead ECG cable with disposable electrodes.
- Electro coutery and defibrillator protraction for OT use.
- Suitable for adult, pediatric & neonatal patients.
- Under Implementation *.
- The Manufactured having ISO certificate & CE Certificate which
is equivalent to US FDA approval.
- Supplied with Three year guarantee.
Other General Features & Specification ECG
- Large 12.1" Color display.
- Approx Dimension: 330mm X 280mm X 150mm.
- Adult to Neonatal Usage.
- Built-in Thermal Printer (Optional).
- Battery Backup.
- User Programmable Alarm setting.
- 24 hour trend information.
ECG
 |
| Lead |
5 leads (R,L,F,N,C) |
| Lead Selection |
1 ; II ; III ; avR ; avL ; avF ; V; MCL 1. |
| Wave Speed (mm/s) |
12.5/ 25 /50. |
| Sensibility |
>200 uV (Peak to peak value). |
| Input Impedance |
> 5 megohm. |
| Heart Rate Range |
0-350 bpm |
| Alarm Range |
0-250 bpm |
| Accuracy |
2% or 2 bpm which ever is greater. |
| Resolution |
1 bpm. |
| Wave Form |
7 Channels. |
| Gain Selection |
X 0.5 ; X 1 ; X 2 ; X 4. |
| Work Status |
OPE, MON, DIA. |
| Lead Fault Alarm |
Audible / Visible. |
| Scale Signal |
1mV (Peak to peak value) |
| Veracity |
3% |
| ST Segment Check |
Measurement Range : -1.0 mV ~ +1.0 mV Alarm Range
: -0.80 mV ~ 0.80 mV |
| ECG Signal Range |
10 (mv) peak to peak values. |
|
 |
Respiration
 |
| Respiration Method |
IRSP. |
| Respiration Rate |
|
| Measurement range |
1 - 99 rpm. |
| Alarm Range |
1 - 99 rpm. |
| Accuracy |
2 rpm / 10% which ever is greater. |
| Respiration Scale |
X0.5 , X1, X 2, X4. |
|
 |
SpO2
 |
| Measurement Range |
0- 100 %. |
| Alarm Range |
0-100 %. |
| Resolution |
1%. |
| Accuracy |
80 - 100 %, 2 % |
| 60 - 79 %, 3 % |
| < 60 %, ignored. |
|
 |
Temperature
 |
| Measurement Range |
0 - 50 C. |
| Alarm Range |
0 - 45 C. |
| Resolution |
0.1 C. |
| Accuracy |
0.2 C. |
|
 |
Thermal Printer (Optional)
 |
| Built In |
|
| Paper Type |
55mm Roll type. |
|
 |
Power Requirement
 |
| AC Input |
220 V 10% 50Hz. |
| Internal Battery |
4 A / Hr 12V lead - Acid
rechargeable. |
| Battery operation time |
60 min. |
| Charge Time |
8 Hours. |
|
 |
NIBP
 |
| Measure Modal |
Auto Surge. |
| Working Mode |
Auto / Manual. |
| Cycle Time |
1 - 60 min. |
| P R Range |
0 - 300 bpm |
| Overload Protection |
Adult / pediatric / neonatal. |
|
 |